Your Questions, Answered

We’ve assembled the most frequently asked questions about Eversense into one place. Got a question that you can’t find the answer to? Contact us.

Filter by topic:

Eversense® E3 Sensor is manufactured by Senseonics across the US and Europe.
Eversense® E3 Transmitter is manufactured by Senseonics in Minnesota, USA.

Yes.
(Customer Service team – see approved talking points/script on account deletion)

Non-clinical testing has demonstrated the Eversense® E3 Sensor is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions:

  • Static magnetic field of 1.5T or 3.0T
  • Maximum spatial field gradient of 2000 gauss/cm (20 T/m).
  • Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 4 W/kg (First Level Controlled Operating Mode).

Under the scan conditions defined above, non-clinical testing results indicate the Eversense Sensor is expected to produce a maximum temperature rise of less than 5.4 °C after 15 minutes of continuous scanning.

In non-clinical testing, the image artifact caused by the device extends approximately 2.83 inches (72 mm) from the Eversense® E3 Sensor when imaged with a gradient echo pulse sequence and a 3T MR system.

The Eversense Smart Transmitter is MR Unsafe and MUST BE REMOVED before undergoing an MRI procedure. Before you undergo an MRI procedure, tell the MRI staff that you have an Eversense Sensor and Smart Transmitter.

X-ray, CT Scan:

DO NOT wear the smart transmitter during medical x-rays or computed tomography (CT) scans. To avoid interference with results, remove the smart transmitter before undergoing medical x-ray or CT scans. Make sure your health care professional knows about your smart transmitter.

Lithotripsy:

The use of lithotripsy is not recommended for people who have an inserted sensor because the effects are unknown.

Diathermy:

DO NOT use diathermy on people who have an inserted sensor. Energy from the diathermy can transfer through the sensor and cause tissue damage in the insertion area.

Electrocautery:

The use of electrocautery near the inserted sensor may damage the device. DO NOT use electrocautery near the sensor.

You should avoid massage therapy near the inserted sensor site. Massage therapy near the sensor site could cause discomfort or skin irritation.

Still have questions?

The Eversense team is dedicated and ready to provide the answers and support you need. Simply click the link below.

The Eversense® E3 Continuous Glucose Monitoring (CGM) System is indicated for continually measuring glucose levels for up to 180 days in persons with diabetes age 18 and older. The system is indicated for use to replace fingerstick blood glucose (BG) measurements for diabetes treatment decisions. Fingerstick BG measurements are still required for calibration primarily one time a day after day 21, and when symptoms do not match CGM information or when taking medications of the tetracycline class. The sensor insertion and removal procedures are performed by a health care provider. The Eversense E3 CGM System is a prescription device; patients should talk to their health care provider to learn more.

The Eversense® 365 Continuous Glucose Monitoring (CGM) System is indicated for continually measuring glucose levels for up to one year in people (18 years and older) with diabetes. The system is indicated for use to replace fingerstick blood glucose (BG) measurements for diabetes treatment decisions. Fingerstick BG measurements are required for calibration one time a week after day 13, and when symptoms do not match CGM information or when taking medications of the tetracycline class. The sensor insertion and removal procedures are performed by a health care provider. The Eversense 365 CGM System is a prescription device; patients should talk to their health care provider to learn more.

For important safety information, see bit.ly/eversensesafety

Eversense, Eversense E3 Continuous Glucose Monitoring, Eversense 365 Continuous Glucose Monitoring, and the Eversense logo are trademarks of Senseonics, Incorporated. Ascensia, the Ascensia Diabetes Care logo are trademarks and/or registered trademarks of Ascensia Diabetes Care Holdings AG. All other trademarks are properties of their respective owners and are used solely for informative purposes. No relationship or endorsement should be inferred or implied.


Apple Watch® is a product of Apple, Inc., and may be separately purchased from an authorized Apple retailer. Apple Watch is not included with the Eversense CGM System. Android is a trademark of Google LLC.

 

PP-SENS-E3-GBL-0060